INSTALLATION, USER AND MAINTENANCE MANUAL FOR WATER OSMOSIS UNIT WO-01
- 36 - 4. Maintenance The osmosis treatment unit runs completely automatically and maintenance is minimal. The only essential operation is periodi
- 37 - Special adjustments: RECIRCULATION Recirculation retreats the waste water to reduce the amount of water rejected. Turn the needle valve “Rec
- 38 - 5. Plumbing diagram
- 39 - LEGEND HYDRAULIC DIAGRAM IN Water inlet F1 Product flow meter E1 Filler solenoid valve P2 Delivery minimum pressure switch P1 Filler p
- 40 - 6. Wiring diagram
- 41 - Electrical specifications: Power supply: 230v AC 50/60Hz Max motor outlet load: 16A_ at 230V AC Max load on solenoid valve 1 outlet: 1A_ a
- 42 - 6- Faults Front door leds The door has two leds to indicate: - Green: power on, normal operation; - Red: fault, production of osmosis-treated
- 43 - B) Osmosis-treated water not produced Possible causes Subcause Solution Display alarm message Check power supply and power switch ON (see
- 44 - C) Water production not conforming Display alarm message Cause of alarm Solution Adjust "bypass" needle valve HHHH The treated
- 45 - Fig.3 Non-return valve Fig.4 Vat cock Fig.5 Pressure switch Solenoid valve Filter
- 46 - 7. Electronic circuit board and programming Hold down F1 and press ENTER to enter conductivity programming. We can now read the
- 47 - In standby (system not running) : Pressing F1 runs a water flushing cycle. X.X indicates the software version (e.g. 2.1) Display: After a
- 48 - Press ENTER or F1: Stop motore se in moto per … (Stop motor after …) The following two messages concern the osmosis pump motor condition. Se
- 49 - Press ENTER or F1; the system returns to standby. The switch at the top of the display, if pressed to the right, switches from the waste w
- 29 - EQUIPMENT FOR POTABLE WATER TREATMENT CONTENTS: Paragraph GENERAL WARNINGS...
- 30 - 1. General warnings STORAGE: THE PACKED APPLIANCE MUST BE STORED IN A DRY, CONDENSATION-FREE AREA, UNDER COVER. THE ADMIT
- 31 - 2. Technical characteristics Dimensions Height 814 mm Empty weight 60 kg Width 376 mm Empty weight including packaging 6
- 32 - Technical characteristics Single phase power voltage 230 v Frequency 50 Hz Osmosis pump power rating 450 w Delivery pump power rating
- 33 - 3. Positioning and installation Positioning: Before installing the appliance, check that there is sufficient clearance for easy ex
- 34 - Water circuit and electrical hookup diagram Installing the membranes Open the front door to access the membrane housings, which must b
- 35 - Pressurizing and starting up the unit: After all pipes have been connected, pressurise the system gradually and check for leaks. Turn the “Os